R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
229
The dialog provides the following additional control elements:
•
Select Cable Type
opens a dialog to use the parameters of a standard (predefined) cable type or
add a cable type with arbitrary parameters.
•
If
Same Offset Line Characteristics at All Ports
is selected, the selected cable type is applied to all
ports.
Remote control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:LOSS<port_no>
[SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet
Mechanical Length
Opens a dialog to define length offset parameters for the physical test ports as mechanical lengths and
permittivities. The dialog also contains the DC loss, the loss at the reference frequency Loss(f
ref
); and the
reference frequency f
ref
; see
Loss parameters: Definition
above.
The
Mechanical Length
dialog contains the three editable columns
Mech. Length, Permittivity
(
]
r
) and
Velocity Fact
1/sqrt(
]
r
)
;
see
Offset parameters: Definition
above. Permittivity and velocity factor are
coupled parameters. The dialog provides the following additional control elements:
•
Select Cable Type
opens a dialog to use the parameters of a standard (predefined) cable type or
add a cable type with arbitrary parameters.
•
If
Same Offset Line Characteristics at All Ports
is selected, the selected cable type is applied to all
ports.
Select
Same Offset Line Characteristics at All Ports
to change the permittivity or velocity factor for all
ports by entering a single value.
Remote control:
[SENSe<Ch>:]CORRection:EDELay<port_no>:DISTance
[SENSe<Ch>:]CORRection:EDELay<port_no>:DIELectric
[SENSe<Ch>:]CORRection:LOSS<port_no>
[SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency
[SENSe<Ch>:]CORRection:LOSS<port_no>:OFFSet