R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
385
as a fixed marker.
CALC:MARK:REF:X 1GHz
Shift the marker horizontally. The response value remains fixed.
CALCulate<Chn>:MARKer<Mk>:REFerence:X <numeric_value>
Defines the stimulus (in Cartesian diagrams: x-axis) value of the reference marker, which can (but doesn't
have to) be displayed using
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON
.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored and may
be set to any value.
<numeric_value>
Stimulus value of the reference marker.
Range [def. unit]
–9.9E+11 Hz to +9.9E+11 Hz [Hz]
*RST value
0 Hz
SCPI, Command
Types
Device-specific, command or query.
Example:
Suppose that the active setup contains an active trace no. 1 and that the
sweep range for a frequency sweep starts at 1 GHz.
CALC:MARK:REF ON
Create the reference marker and display it in the center of the sweep range.
CALC:MARK:REF:X 1GHz
Set the reference marker to the beginning of the sweep range.
CALCulate<Chn>:MARKer<Mk>:REFerence:Y?
Returns the response (in Cartesian diagrams: y-axis) value of the reference marker. The reference marker
must be created before using
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON
.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored and may be
set to any value.
Response
Response value of the reference marker.
Range [def. unit]
Depending on the measured quantity. [unit depending on the marker format; see
CALCulate<Chn>:MARKer<Mk>:FORMat
]
*RST value
–
SCPI,
Command
Types
Device-specific, query only.
Example:
Suppose that the active setup contains an active trace no. 1.
CALC:MARK:REF ON