![Rohde & Schwarz ZVL series Operating Manual Download Page 23](http://html.mh-extra.com/html/rohde-and-schwarz/zvl-series/zvl-series_operating-manual_1477904023.webp)
R&S
ZVL
Glossary of Terms
Quick Start Guide 1303.6538.62 - 05
7
error can be corrected by means of a full one-port calibration or a two-port
calibration (except normalization).
Stimulus value:
Value of the sweep variable (frequency) where a measurement is
taken. Also termed sweep point.
Sweep:
Series of consecutive measurements taken at a specified sequence of
stimulus values = series of consecutive measurement points.
Sweep point:
Value of the sweep variable (stimulus value: frequency) where a
measurement is taken.
Sweep range:
Continuous range of the sweep variable (frequency) containing the
sweep points where the analyzer takes measurements. In a Segmented
Frequency sweep the sweep range can be composed of several parameter
ranges or single points.
Sweep segment:
Continuous frequency range or single frequency point where the
analyzer measures at specified instrument settings (generator power, IF
bandwidth etc.). In the Segmented Frequency sweep type the entire sweep
range can be composed of several sweep segments.
T
TOSM:
A calibration type using four known standards (through, open, short, match),
also called SOLT or 12-term error correction model.
Trace:
A trace is a set of data points that can be displayed together on the screen. The
trace settings specify the mathematical operations used to obtain traces from
the collected data. They complement the definitions of the Channel menu.
Each trace is assigned to a channel. The channel settings apply to all traces
assigned to the channel.
Trace point:
Point on the screen which is an element of the displayed trace.
Transmission tracking error:
Frequency-dependent variation of the ratio of the
transmitted wave to the reference wave at a test port when an ideal
transmission coefficient (= 1) is measured. The transmission tracking error can
be corrected by means of a transmission normalization or one of the more
sophisticated calibration methods.
W
Window:
Rectangular portion of the screen showing all diagram areas of a particular
setup. Windows are limited by a blue frame with several icons. The analyzer
uses standard windows provided by the operating system.