R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
400
Use
CALCulate<Chn>:RIPPle:FAIL?
to query the result for global ripple limit check.
<Chn>
Channel number used to identify the active trace.
<Seg>
Ripple limit range number
Response:
<Fail>,
<Limit>
0 | 1 - 0 represents pass, 1 represents fail
Measured ripple in the limit range. A result is returned even if the limit check in
the range no. <Seg> is disabled; see example below.
*RST value
– (a reset deletes all ripple limit ranges)
SCPI, Command
Types
Device-specific, query only
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,3
Define and enable a ripple limit range in the stimulus range between 1.5 GHz
and 2 GHz, assigning a ripple limit of +3 dBm.
CALC:RIPP:STAT ON; SEGM:RES?
Enable the limit check and query the result for the created range. Possible
response: 0,0.3529814004.
CALC:RIPP:DATA 0,2500000000, 3000000000,3
Define a second ripple limit range with disabled limit check (no limit check
results are displayed in the diagram area).
CALC:RIPP:SEGM2:RES?
Query the result for the second range. Possible response:
0,1.149071925
.
CALCulate<Chn>:RIPPle:SEGMent<Seg>:STIMulus:STARt <start_value>
Changes the start stimulus value (i.e. the smallest stimulus value) of a ripple limit range. A range must be
created first to enable this command (e.g
CALCulate<Chn>:RIPPle:DATA
).
To define several ripple limit ranges with a single command, use
CALCulate<Chn>:RIPPle:DATA
.
<Chn>
Channel number used to identify the active trace.
<Seg>
Ripple limit range number
<start_value>
Frequency value.
Range [def. unit]
Almost no restriction for limit ranges; see
Rules for Limit Line Definition.
[Hz. For distance-to-fault traces the default unit is the length unit defined via
UNIT:LENGth
.]
*RST value
– (no ripple limit line defined after *RST)
SCPI, Command
Types
Device-specific, with query.
Example:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,3
Define and enable a ripple limit range in the stimulus range between 1.5 GHz
and 2 GHz, assigning a ripple limit of +3 dBm.
CALC:RIPP:SEGM:STIM:STAR 1GHZ; STOP 2.5 GHZ; :CALC:RIPP:SEGM:LIM 5
Change the range to a stimulus range between 1 GHz and 2.5 GHz and a limit