R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
554
TRIGger<Ch>...
This subsystem is used to synchronize the analyzer's actions (in particular, the start of a measurement
sequence) with events.
TRIGger<Ch>[:SEQuence]:HOLDoff <numeric_value>
Sets a delay time between the trigger event and the start of the measurement
(Trigger Delay)
.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
<numeric_value>
Delay time.
Range [def. unit]
0 s to 13680 s [s]. UP and DOWN increment/decrement the trigger delay in
10 ms steps.
*RST value
0 s
SCPI, Command
Types
Confirmed, command or query
Example:
TRIG:SOUR TIM
Activate internal timer as trigger source.
TRIG:HOLD UP
Set a delay time of 10 ms.
TRIGger<Ch>[:SEQuence]:LINK 'POINt' | 'SWEep'
Selects the
Triggered Meas. Sequence
. The identifier for the sequence is a string variable.
<Ch>
Channel number.
'SWEep' |
Trigger event starts an entire sweep.
'POINt' |
Trigger event starts measurement at the next sweep point. This mode is not
available on R&S ZVL13 network analyzers.
*RST value
'SWEep'
SCPI, Command
Types
Confirmed, command or query
Example:
TRIG:LINK 'POIN'
Select a sweep point as triggered measurement sequence.
TRIGger<Ch>[:SEQuence]:SLOPe POSitive | NEGative
Qualifies whether the trigger event occurs on the rising or on the falling edge of the external TTL trigger
signal.
<Ch>
Channel number.
POSitive |
Rising edge.
NEGative
Falling edge.