R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
223
Checks on switching to the next dialog
When the
Next>
buttonis pressed the analyzer checks the calibration kits and the matching of the
calibration standards and possibly displays a notice box (confirm with
OK).
This happens if one of the
calibration kits is described by ideal kit parameters or typical values.
Remote control:
[SENSe<Ch>:]CORRection:COLLect:CONNection<port_no>
[SENSe<Ch>:]CORRection:COLLect:SCONnection<port_no>
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:SELect "<Ckit_Name>"
Measure Standards
The last dialog of the calibration wizard is used to perform the necessary measurements of standards and
to calculate the correction data.
The dialog displays the list of measured standards compiled in the previous dialog. The list is complete: All
standards must be measured to perform the selected calibrations.
Structure of the Measured Standards list
The list of measured standards has a tree structure.
•
The first level contains all physical ports where one-port
(reflection) measurements are
required and all physical port combinations where two-port
(transmission)
measurements must be performed.
•
The second level contains check boxes for the standards to be measured at each port or
port combination.
•
For a sliding match, a third level contains check boxes for the different positions of the
sliding element.
To expand the list of a physical port symbol, click the symbol, to collapse the list, double-click
the symbol.
Calibrations using a match or sliding match
If the calibration kit contains a sliding match standard, the
Sliding Match
appears in the
Measured