
R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
222
The table contains the following rows:
•
Physical Port Number #
The ports (and therefore the number of table rows) are determined by the active calibration type
selected in the
Start Cal
submenu.
•
Connector
provides a drop-down list to select the connector type An (f) behind the connector type denotes
female connectors, an (m) denotes male connectors. Symmetric (sexless) connectors (e.g. PC7)
are not labeled. User-defined connectors can be added or removed in the Available Connector
Types dialog, which is opened from the Channel – Cal – Cal Kits dialog. at the port and its gender.
If
Same Connector at All Ports in Table
is active, the connector types at all ports (but not their
gender) are always adjusted to the current selection. The default selection depends on the
analyzer model, e.g. on the reference impedance (50
V
/ 75
V
).
•
Ref Imp
shows the reference impedance for the selected connector.
•
Calibration Kits
provides a drop-down list to select a calibration kit. The list contains all calibration kits available for
the selected connector type. The assignment of a calibration kit to a connector type must be the
same for all physical ports: If a calibration kit is changed, the analyzer automatically assigns the
new kit to all ports with the same connector type.
Completness:
The selected calibration kit must contain all standards needed for the active
calibration type. If it doesn't the analyzer displays an error message.
Waveguide cutoff:
If a user-defined waveguide connector is assigned to one of the calibrated
ports, then the start frequency of the active channel must be above the waveguide cutoff.
Otherwise the analyzer displays an error message.
Import Kit
opens the
Import Calibration Kit
dialog to load and (if desired) activate a cal kit file.
Next >
opens the second dialog of the wizard to continue the calibration procedure.
If the calibrated channel has already been assigned to a cal group, the correction data overwrites the
cal group data, so the new calibration will affect all channels assigned to the cal group. The
network analyzer generates a notice message "New calibration will overwrite cal pool!" when
opening the calibration wizard.