R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
452
Create channel no. 2 with a new trace and start a single sweep in channel no. 2.
INITiate<Ch>[:IMMediate]
Starts a new single sweep sequence. This command is available in single sweep mode only
(INITiate<Ch>:CONTinuous OFF). The data of the last sweep (or previous sweeps, see
Sweep History)
can be read using CALCulate<Ch>:DATA:NSWeep? SDATa, <history_count>.
In contrast to all other commands of the analyzer,
INITiate<Ch>[:IMMediate]
has been
implemented for overlapped execution; see Command Synchronization.
<Ch>
Channel number. If the channel number does not exist the analyzer returns an error
message. If
INITiate<Ch>[:IMMediate]:SCOPe ALL
is active, this suffix is
ignored.
*RST value
–
SCPI,
Command
Types
Confirmed, no query
Example:
See INITiate<Ch>:CONTinuous
INITiate<Ch>[:IMMediate]:SCOPe ALL | SINGle
Selects the scope of the single sweep sequence. The setting is applied in single sweep mode only
(INITiate<Ch>:CONTinuous OFF).
<Ch>
Channel number. The channel is ignored because the command affects all
channels.
ALL
INITiate<Ch>[:IMMediate] starts a single sweep in all channels.
SINGle
INITiate<Ch>[:IMMediate] starts a single sweep in the referenced channel
<Ch> only.
*RST value
ALL
SCPI, Command
Types
Device-specific, command or query
Example:
See INITiate<Ch>:CONTinuous
INPut...
This subsystem controls the characteristics of the analyzer’s input ports.
INPut<port_no>:ATTenuation <numeric_value>
Sets the attenuation for the received waves.
Note:
INPut<port_no>:ATTenuation
is not channel-specific; the value is valid for all channels. Use