R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
370
CALCulate<Chn>:LIMit:SEGMent<Seg>:TYPE LMIN | LMAX | OFF
Selects the limit line type for a limit line segment. This can be done before or after defining the stimulus
and response values of the segment, however, a segment must be created first to enable this command
(e.g
CALC:LIM:DATA
).
The type command overwrites the CALCulate<Chn>:LIMit:DATA settings and is overwritten by them.
It is not affected by the other commands in the LIMit subsystem defining stimulus and response
values of limit lines.
<Chn>
Channel number used to identify the active trace.
<Seg>
Segment number
Parameters
Limit line type
Range [def. unit]
LMAX (upper limit line segment), LMIN (lower limit line segment), OFF (limit
check switched off, limit line segment not deleted)
[–]
*RST value
LMAX
SCPI, Command
Types
Device-specific, with query.
Example:
*RST; CALC:LIM:UPP 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between 1 GHz and 2
GHz, using default response values.
CALC:LIM:SEG:TYPE LMIN
Turn the defined limit line segment into a lower limit line segment.
CALCulate<Chn>:LIMit:STATe
Switches the limit check (including upper and lower limits) on or off.
Use
CALCulate<Chn>:LIMit:UPPer:STATe
or
CALCulate<Chn>:LIMit:LOWer:STATe
to
switch on or off the individual limit checks for upper or lower limit lines.
<Chn>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Limit check on or off.
*RST value
OFF
SCPI, Command
Types
Confirmed, command or query.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between 1 GHz and 2
GHz, using default response values.
CALC:LIM:STAT ON; CALC:LIM:FAIL?
Switch the limit check on and query the result.