R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
384
*RST value
CONT
SCPI, Command
Types
Device-specific, command or query.
Example:
Suppose that the active setup contains an active trace no. 1.
CALC:MARK:REF:MODE DISC; :CALC:MARK2:REF:MODE CONT
Create the reference marker in discrete mode and marker 2 in continuous
mode.
CALC:MARK:REF ON; :CALC:MARK2 ON
Display the two markers. Due to the different modes the horizontal position can
be different.
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] <Boolean>
Creates the reference marker and assigns it to trace no. <Chn>.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored and may be
set to any value.
<Boolean>
ON | OFF - Creates or removes the marker.
*RST value
OFF
SCPI, Command
Types
Device-specific, command or query.
Example:
Suppose that the active setup contains an active trace no. 1.
CALC:MARK:REF ON; CALC:MARK ON
Create the reference marker and marker 1 and assign them to trace no. 1. The
default position of both markers is the center of the sweep range.
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE NORMal | FIXed
Sets the reference to normal or fixed mode. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON
.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored and may
be set to any value.
Parameters
NORMal – response value changes according to the measurement result.
FIXed – marker keeps its current response value.
*RST value
NORMal
SCPI, Command
Types
Device-specific, command or query.
Example:
CALC:MARK:REF ON; CALC:MARK:REF:TYPE FIX
Create the reference markerand display it in the center of the sweep range