R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
403
CALCulate<Chn>:STATistics...
This subsystem evaluates and displays statistical and phase information of the trace.
CALCulate<Chn>:STATistics:DOMain:USER <numeric_value>
Selects one out of 10 evaluation ranges to be configured with the
CALCulate<Chn>:STATistics:DOMain:USER:STARt
and
CALCulate<Chn>:STATistics:DOMain:USER:STOP
commands.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
<numeric_value>
Number of the evaluation range.
Range [def. unit]
1 to 10. In addition, 0 denotes the (non-configurable)
Full Span
evaluation
range. [–]
*RST value
0
SCPI, Command
Types
Device-specific, command or query.
Example:
*RST; CALC:STAT:DOM:USER?
Query the default evaluation range. The response is zero, i.e. the evaluation
range is equal to the complete sweep range
CALC:STAT:DOM:USER 1; :CALC:STAT:DOM:USER:STARt 1GHZ; STOP 2GHZ
Select evaluation range no. 1 and define the evaluation range between 1
GHz and 2 GHz.
CALCulate<Chn>:STATistics:DOMain:USER:STARt <numeric_value>
Defines the start value of the evaluation range selected via
CALCulate<Chn>:STATistics:DOMain:USER
.
<Chn>
Channel number used to identify the active trace.
<numeric_value>
Start value of the evaluation range.
Range [def. unit]
–1000 GHz to 999.999999999 GHz. [Hz]
*RST value
Lowest frequency of the analyzer, depending on the analyzer model.
SCPI, Command Types
Device-specific, command or query.
Example:
See CALCulate<Chn>:STATistics:DOMain:USER
CALCulate<Chn>:STATistics:DOMain:USER:STOP <numeric_value>
Defines the stop value of the evaluation range selected via
CALCulate<Chn>:STATistics:DOMain:USER
.
<Chn>
Channel number used to identify the active trace.
<numeric_value>
Stop value of the evaluation range.