R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
442
Types
Example:
See DISPlay:WINDow<Wnd>:TRACe<WndTr>:Y[:SCALe]:BOTTom
Units for DISPlay... Commands
The DISPlay... subsystem contains commands to define particular points in the diagram, e.g. to set the
scale or a reference value. This requires the entry of a numeric value and a physical unit, depending on
the parameter type displayed. The following table lists the physical units accepted by the analyzer.
Power
DBM, DB, DBW, W, MW, UW, NW, PW
Voltage
V, MV, UV, NV, PV, DBV, DBMV, DBUV
Phase
DEG, KDEG, MDEG, UDEG, NDEG, PDEG
Group delay
S, MS, US, NS, PS
Impedance
OHM, GOHM, MOHM, KOHM
Admittance
SIE, MSIE, USIE, NSIE
Inductance
H, MH, UH, NH, PH, FH
Capacitance
F, MF, UF, NF, PF, FF
Dimensionless
UNIT, MUNIT, UUNIT, NUNIT, PUNIT, FUNIT
FORMat...
This subsystem sets a data format for transferring numeric and array information from and to the analyzer.
FORMat:BORDer NORMal | SWAPped
Controls whether binary data is transferred in normal or swapped byte order.
Parameters
SWAPped – The least significant bit is transferred first (little endian)
NORMal – The most significant bit is transferred first (big endian)
*RST value
SWAPpped (if the
GPIB Language
is set to
PNA
or
HP xxxx,
then the order
is NORMal)
SCPI, Command
Types
Confirmed, command or query
Example:
FORM:BORD NORM
Change the bit order to normal mode.