R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
391
Command
Types
Example:
*RST; CALC:MATH:MEM
Copy the current state of the default trace
Trc1
to a memory trace named
'Mem2[Trc1]'. The memory trace is not displayed.
CALC:MATH:FUNC DIV
Define a mathematical trace, dividing the data trace by the stored memory trace.
The mathematical trace is displayed instead of the active data trace.
CALC:MATH:STAT?
The response is 1 (mathematical mode switched on, mathematical trace
displayed).
CALCulate<Chn>:MATH:MEMorize
Copies the current state of the active data trace to a memory trace. If a mathematical trace is active, the
data trace associated with the mathematical trace is copied. The memory trace is named
Mem<n>[<Data_Trace>]
where <n> counts all data and memory traces in the active setup in chronological
order, and
<Data_Trace>
is the name of the associated (copied) data trace.
The exact function of the command depends on the number of memory traces associated with the active
data trace:
•
If no memory trace is associated to the active trace, a new memory trace is generated.
•
If several memory traces are associated with the active trace, the current measurement data
overwrites the last generated or changed memory trace.
To copy a trace to the memory without overwriting an existing memory trace or define a memory trace
name, use TRACe:COPY <memory_trc>,<data_trc>. To copy an active mathematical trace use
TRACe:COPY:MATH <memory_trc>,<data_trc>
<Chn>
Channel number used to identify the active trace.
*RST value
–
SCPI, Command
Types
Device-specific, no query
Example:
*RST; CALC:MATH:MEM
Copy the current state of the default trace
Trc1
to a memory trace named
'Mem2[Trc1]'. The memory trace is not displayed.
DISP:WIND:TRAC2:FEED 'Mem2[Trc1]'
Display the created memory trace in the active diagram area (diagram area no.
1).
CALCulate<Ch>:PARameter...
This subsystem assigns names and measurement parameters to traces. The commands are device-
specific.