R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
527
Disable the measurement in the created sweep segment.
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl <Meas. Delay>
Defines the delay time for each partial measurement in sweep segment no. <Seg>
(Meas. Delay
). If
coupling of the segments is switched on (
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol ON
)
the delay is valid for all sweep segments in the current channel. This command is not available on R&S
ZVL13 network analyzers.
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Meas.
Delay>
Meas. delay before each partial measurement.
Range [def.
unit]
0 s to 2.5E+003 s. [s] Changing the delay leaves the number of points unchanged
but has an impact on the duration of the sweep
([SENSe<Ch>:]SEGMent<Seg>:SWEep:TIME).
*RST value
0 s
SCPI,
Command
Types
Device-specific, command or query. This default value corresponds to automatic
sweep time setting in manual control.
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings.
SEGM:SWE:DWEL 1 MS
Set the meas. delay in segment no. 1 to 1 ms.
SEGM:DEF? Response: 300000,8000000000,51,-300,0.056559,0,10000,POS,NORM
Query the channel parameters for sweep segment 1. The response value for the
segment sweep time (olive) implicitly contains the defined meas. delay.
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol <Boolean>
Qualifies whether or not the
Meas. Delay
defined via
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
can
be set independently for each sweep segment.
<Ch>
Channel number.
<Seg>
Sweep segment number. This suffix is ignored; the command controls the whole
segmented sweep.
<Boolean>
ON: The meas. delay can be set independently for each sweep segment.
OFF: The meas. delay in all sweep segments is equal to the
Meas. Delay
measurements for unsegmented sweeps set via
[SENSe<Ch>:]SWEep:DWELl
.
*RST value
OFF. The parameter is automatically switched to ON when a meas. delay time is
entered using
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl
.
SCPI,
Command
Types
Device-specific, command or query.