R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
489
<Min_Freq>
Min. frequency for which the circuit model is valid
Default unit is Hz
<Max_Freq>
Max. frequency for which the circuit model is valid
Default unit is Hz
<El_Length>
Electrical length (offset parameter) of the standard
Default unit is m or ft,
depending on the
Distance
Unit
setting in the
System
Configuration
dialog
<Loss>
Loss (offset parameter) of the standard
To be specified without
unit (implicit unit is dB)
<Z0>
Reference impedance (no unit)
To be specified without
unit (implicit unit is
V
)
<C0> | <L0>
Polynomial coefficients C
0
or L
0
for the fringing capacitance or residual
inductance of the standard (load parameter)
To be specified without
unit (implicit unit is fF or
pH)
<C1> | <L1>
Polynomial coefficients C
1
or L
1
for the fringing capacitance or residual
inductance of the standard (load parameter)
To be specified without
unit (implicit unit is fF /GHz
or pH /GHz)
<C2> | <L2>
Polynomial coefficients C
2
or L
2
for the fringing capacitance or residual
inductance of the standard (load parameter)
To be specified without
unit (implicit unit is
fF/GHz
2
or pH/GHz
2
)
<C3> | <L3>
Polynomial coefficients C
3
or L
3
for the fringing capacitance or residual
inductance of the standard (load parameter)
To be specified without
unit (implicit unit is
fF/GHz
3
or pH/GHz
3
)
OPEN | SHORt |
MATCh
Simplified modelling as an open or short or match standard. The load
circuit model generally consists of capacitance C which is connected in
parallel to an inductance L and a resistance R, both connected in series.
OPEN means that R is infinite so that the standard behaves like a
capacitor (no inductance, the polynomial coefficients specify C). SHORt
means that R is zero so that the standard behaves like an inductance (no
capacitance, the polynomial coefficients specify L). MATCh means that
the standard behaves like a match with the specified <Resistance>
Character data
<Resistance>
Optional resistance for the MATCh standard
Numeric value
, <Port_1>,
<Port_2>
Optional port restriction: one port number for one port standards, two port
numbers for two port standards
Integer value
The different standard types are defined by the following parameters:
<std_type> Meaning
Parameters
MOPen | FOPen
Open (m) or open (f)
'<Ckit_Name>' ... <C3>[, <Port_1>]
complete parameter list with capacitance
coefficients, no OPEN | SHORT | MATCh
MSHort | FSHort
Short (m) or short (f)
'<Ckit_Name>' ... <L3>[, <Port_1>]
complete parameter list with inductance