R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
432
This mode is recommended for preserving the display especially if the instrument is exclusively
operated via remote control.
<Boolean>
ON | OFF - Switching the power-save mode on or off.
*RST value
OFF
SCPI, Command Types
Device-specific, command or query
Example:
*RST; DISP:PSAVe ON
Switches on the powersave mode.
DISPlay[:WINDow<Wnd>]:MAXimize <Boolean>
Maximizes all diagram areas in the active setup or restores the previous display configuration.
<Wnd>
Number of the diagram area to become the active diagram area.
DISPlay:WINDow<Wnd>:MAXimize
acts on all diagrams of the current setup,
however, the diagram no. <Wnd> is displayed on top of the others.
<Boolean>
ON | OFF - Maximize all diagram areas or restore the previous display
configuration.
*RST value
OFF
SCPI,
Command
Types
Device-specific, command or query (returns whether or not the diagrams are
maximized).
Example:
*RST; DISP:WIND2:STAT ON
Create diagram areas no. 1 (with default trace) and 2 (with no trace).
DISP:WIND2:MAXimize ON
Maximize the diagram areas, placing area no. 2 on top.
DISPlay[:WINDow<Wnd>]:STATe <Boolean>
Creates or deletes a diagram area, identified by its area number <Wnd>.
<Wnd>
Number of the diagram area to be created or deleted.
<Boolean>
ON | OFF - Creates or deletes diagram area no. <Wnd>.
*RST value
–
SCPI, Command
Types
Device-specific, command or query (returns whether or not a particular
diagram area exists).
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the input reflection
coefficient S11.
DISP:WIND2:STAT ON