R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
431
Assign the S-parameter S
11
to the default trace, open the local analyzer screen,
then display the phase of the measured quantity. The commands are executed
immediately without any manual entry.
DISP:MENU:KEY:EXEC 'Start'
Open the numeric input bar for the start frequency of the sweep. The frequency
can be entered manually.
DISP:MENU:KEY:EXEC 'About Nwa'
Display information about your network analyzer and the firmware version. The
info box is closed when you click the analyzer screen or if you send another
command.
DISPlay:MENU:KEY:SELect '<menu_key>'
Activates the menu or submenu of the specified key.
'<menu_key>'
Name of the key as shown in the analyzer's softkey bar (case-sensitive
string variable).
*RST value
–
SCPI, Command
Types
Device-specific, no query
Example:
*RST; DISP:MENU:KEY:SEL 'S11'
Open the
Trace
–
Meas
menu in order to select a measured quantity.
DISPlay:PSAVe:HOLDoff <holdoff time>
Sets the holdoff time for the powersave mode of the display.
<holdoff time>
Holdoff time for the powersave mode.
Range [def. unit]
1 to 60 minutes, the resolution is 1 minute. The entry is dimensionless.
*RST value
15
SCPI, Command Types
Device-specific, command or query
Example:
*RST; DISP:PSAV:HOLD 30
Sets the holdoff time to 30 minutes.
DISPlay:PSAVe[:STATe] <Boolean>
This command switches on or off the powersave mode of the display. With the powersave mode activated
the display including backlight is completely switched off after the elapse of the response time (see
DISPlay:PSAVe:HOLDoff command).