R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
492
[SENSe<Ch>:]CORRection:CKIT:INSTall '<file_name>'
Loads cal kit data from a specified ZVR cal kit file.
<Ch>
Channel number. This suffix is ignored because calibration kits are channel-
independent.
'<file_name>'
String parameter to specify the name and directory of the cal kit file to be loaded.
Note: The loaded file must be a ZVR-specific cal kit file with the extension *.ck.
NWA cal kit files (*.calkit) can be imported using the
MMEMory:LOAD:CKIT
command. Agilent cal kit files can be imported manually and converted into
*.calkit files.
*RST value
–
SCPI,
Command
Types
Device-specific, no query.
Example:
CORR:CKIT:INST
'C:\R_S\Instr\user\Nwa\Calibration\Kits\ZCAN.ck'
Load the previously created ZVR cal kit file ZCAN.ck from the
default cal kit directory.
MMEM:STOR:CKIT 'ZCAN',
'C:\R_S\Instr\user\Nwa\Calibration\Kits\ZCAN.calkit'
Store the imported cal kit data to a NWA cal kit file
ZCAN.calkit (assuming that the cal kit name stored in ZCAN.ck
reads ZCAN).
[SENSe<Ch>:]CORRection:COLLect...
This subsystem controls the system error correction and manages calibration kits.
[SENSe<Ch>:]CORRection:COLLect[:ACQuire] THRough | OPEN1 | OPEN2 |
OPEN12 | SHORT1 | SHORT2 | SHORT12 | MATCH1 | MATCH2 | MATCH12 | NET |
ATT | IMATCH12 | REFL1 | REFL2 | SLIDe1 | SLIDE2 | SLIDE12 | LINE1 | LINE2 |
M1O2 | O1M2 | M1S2 | S1M2 | OSHORT1 | OSHORT2
Starts a calibration measurement in order to acquire measurement data for the selected standards. The
standards are reflection or transmission standards and must be connected to port 1 or 2 of the analyzer.
Use the generalized command
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
to
obtain measurement data at arbitrary analyzer ports.
<Ch>
Channel number of the calibrated channel. If unspecified the numeric suffix is set to 1.
THRough ...
OSHORT2
Standard types:
Through, Open, Short, Match (MATCH12 and IMATCH12 are
synonymous), Symmetric Network (NET), Attenuation (ATT), Reflect, Sliding Match
(SLIDe), Line1 (LINE1 and LINE are synonymous), Line2, Match/Open (M1O2,
O1M2), Match/Short (M1S2, S1M2), Offset Short (OSHort)
.
The numbers in the parameters denote the analyzer ports. Two numbers 12 mean that