R&S ZVL
GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
131
well-defined.
Remote control:
CALCulate<Chn>:MATH:FUNCtion SUBTract
Show Data
Displays or hides the active data trace in the diagram area. If the mathematical option
Math = Data/Mem
or
Math = Data-Mem
is active, then the active mathematical trace is displayed or hidden.
Remote control:
No command, display configuration only.
Show Mem
Displays or hides the active memory trace in the diagram area or the memory trace associated with the
active data trace.
If no memory trace is associated with the active data trace,
Show Mem
is disabled.
Remote control:
No command, display configuration only.
Traces
The
Traces
submenu provides functions to handle traces and diagram areas, and assign traces to
channels.
•
Select Trace
opens a box to select an arbitrary trace of the active setup as the active trace
(disabled if only one trace is defined).
•
Add Trace
creates a new trace in the current diagram area.
•
Add Trace + Diag. Area
generates a new trace in a new diagram area.
•
Delete Trace
deletes the active trace.
•
Assign Diag. Area
assigns the active trace to another diagram area.
•
Assign Channel
assigns the active trace to another channel.
•
Trace Manager
opens a dialog to perform the previous actions systematically for all traces and
diagram areas.