R&S ZVL
Programming Examples
Basic Tasks
Operating Manual 1303.6580.32-05
570
// The response is 'Trc1,S21,Impedance_trace,Z-S21,Admittance_trace,Y-S21'
//
// Query the reference level for the 'Z_trace'.
// The trace is referenced by its number in diagram area no. 2
.
DISPlay:WINDow2:TRACe3:Y:RLEVel?
//
// Change the display format for the 'Z_trace'. The trace is the active trace in channel 3,
// so it is referenced by the channel suffix 3
.
CALCulate3:FORMat PHASe
Markers and Limit Lines...
Programming task:
Display two traces in a single diagram ares, use markers to read results, and perform
a limit check.
Important remote control features for this program example
The following command sequence illustrates the structure of the remote commands discussed in section
Basic Remote Control Concepts.
In particular it shows that:
Traces are referenced by trace names. The active trace of a channel is often referenced by the
channel suffix. This simplifies the program syntax, e.g. in the commands for marker settings and
for the limit check.