![Rohde & Schwarz ZVL series Operating Manual Download Page 321](http://html.mh-extra.com/html/rohde-and-schwarz/zvl-series/zvl-series_operating-manual_1477904321.webp)
R&S ZVL
GUI Reference
MODE Key
Operating Manual 1303.6580.32-06
306
last saving process is counted up to the next unused name. If you use a file name that already exists,
upon saving, a message is displayed. Selecting Yes overwrites the existing file, selecting No aborts the
saving process. For further details on the file name and an example, refer to the file menu,
Edit File Name
softkey.
Remote Control:
HCOP:ITEM:ALL
HCOP
HCOP:NEXT
Device Setup
Opens the
Device Setup
dialog box to select file format and the printer. The dialog box consists of two
tabs which are selected via the
Device 1/2
softkey.
Remote Control:
HCOP:DEV:LANG GDI
SYST:COMM:PRIN:ENUM:FIRS?
SYST:COMM:PRIN:ENUM:NEXT?
SYST:COMM:PRIN:SEL <Printer>
HCOP:PAGE:ORI PORT
HCOP:DEST "SYST:COMM:PRIN"
Device 1/2
Selects the tab of the device in the
Device Setup
dialog box. The analyzer is able to manage two print
settings independently of each other. For each device the print setting is displayed on the corresponding
tab of the
Device Setup
dialog box (
Device Setup
softkey).
Comment
Opens dialog box to enter a comment. Max. 120 characters are allowed. 60 characters fit in one line. In
the first line, at any point a manual line–feed can be forced by entering "@".
Date and time are inserted automatically. The comment is printed below the diagram area, but not
displayed on the screen. If a comment should not be printed, it must be deleted.
Install Printer
Opens the
Printers and Faxes
window to install a new printer. All printers that are already installed are
displayed.
With firmware versions V3.00 and higher, a user account with administrator rights is necessary to install a
printer (see
Operation with and without Administrator Rights
).
MODE Key
The MODE key selects the operating mode of the analyzer. The operating mode is stored when the