R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
260
Define Restart
Opens a dialog to configure the
Single
sweep mode.
•
Number of Sweeps
selects the number of sweeps to be measured in single sweep mode: either
one
(Single Sweep)
or a group of consecutive sweeps.
•
Scope
defines whether the active sweep mode (single or continuous) and the
Number of Sweeps
are valid in the
Active Channel
only or for
All Channels
in the active setup. If
All Channels
is
selected, the number of sweeps in the single sweep sequence is equal to the selected
Number of
Sweeps
times the number of channels. The sequence starts with the first sweep in channel no. 1.
In remote control, it is possible to retrieve the results acquired in any of the sweeps within a single
sweep group; see
Sweep History.
Remote control:
[SENSe<Ch>:]SWEep:COUNt
INITiate<Ch>:IMMediate:SCOPe
Trigger
The
Trigger
submenu selects the source of the trigger signal and provides additional
Trigger Settings
.
Trigger system of the analyzer
The trigger system is used to synchronize the analyzer's actions with events that can be provided by an
external signal. Triggered measurements are an alternative to the default mode
(Free Run, Continuous
Sweep)
, where the measurement is continuously repeated without fixed time reference.
A trigger event may start an entire sweep or a single sweep point. Moreover, it is possible to define a
delay between trigger events and the measurement sequences.
The trigger settings are also valid for calibration sweeps (see Measure Standards dialog). This means
that, in external trigger mode, the external trigger signal must be available during the system error
correction, too. To start the calibration sweeps without delay, use the Free Run trigger type.