R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
392
CALCulate<Ch>:PARameter:CATalog?
Returns the trace names and measurement parameters of all traces assigned to a particular channel.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
Response
String parameter with comma-separated list of trace names and measurement
parameters, e.g.
'CH4TR1,S11,CH4TR2,S12'
. The measurement parameters are
returned according to the naming convention of
CALCulate<Ch>:PARameter:SDEFine. The order of traces in the list reflects their
creation time: The oldest trace is the first, the newest trace is the last trace in the list.
*RST value
–
SCPI,
Command
Types
Device-specific, query only.
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the input reflection
coefficient S11.
CALC4:PAR:CAT?
Query the traces assigned to channel 4. If
Ch4Tr1
is the only trace assigned to
channel 4, the response is
'CH4TR1,S11'.
CALCulate<Ch>:PARameter:DEFine '<string>', S11 | S12 | S21 | S22
Creates a trace and assigns a channel number, a name and a measurement parameter to it. The trace is
not displayed. To display a trace defined via
CALCulate<Ch>:PARameter:DEFine
, a window must be
created (
DISPlay:WINDow<Wnd>[:STATe] ON
) and the trace must be assigned to this window
(
DISPlay:WINDow<Wnd>:TRACe:FEED
); see example below.
Traces must be selected to become active traces; see
CALCulate:PARameter:SELect
.
This command has been implemented for compatibility reasons. The parameter names in this
command differ from ZVL conventions; moreover the parameter list is not complete. The
alternative command CALCulate<Ch>:PARameter:SDEFine uses a complete parameter list with
compatible names.
<Ch>
Channel number. <Ch> may be used to reference a previously defined channel. If
<Ch> does not exist, it is generated with default channel settings.
'<string>'
Trace name, e.g. 'Trc4'. See
Rules for trace names
in the
Trace Manager
description.
S11, S12, S21,
S22
Measurement parameter
*RST value
–
SCPI,
Command
Types
Device-specific, no query.
CALCulate<Ch>:PARameter:CATalog?
returns a list
of all defined traces.
Example:
CALC4:PAR:DEF 'Ch4Tr1', S11