R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
503
types.
*RST value
'N 50 Ohm',FEM for all ports.
SCPI,
Command
Types
Device-specific, command or query.
Example:
*RST; CORR:COLL:SCON1 'N 75 Ohm', MALE; SCON4?
Change the connector type at port 1 from '
N 50 Ohm', FEM
to '
N 75 Ohm', MALE
.
The connector type at the other ports is also changed to N 75 Ohm, however, the
gender (female) is maintained.
CORR:COLL:SCON4?
returns
'N 75 Ohm', FEM
.
[SENSe<Ch>:]CORRection:CONNection '<conn_name>', TEM | WGUide, GENDer
| NGENder, <perm_rel>, <imped>
Configures the user-defined connector types.
<Ch>
Channel number
'<conn_name>'
Name of the user-defined connectors, string parameter
TEM | WGUide
Transverse electric or waveguide type propagation mode
GENDer | NGENder
Polar connector type (m/f)
Sexless connector type
<perm_rel>
Relative permittivity
Range [def. unit]
0.0000000001 to 1000. [–] UP/DOWN/MIN/MAX parameters are not
available for this command
*RST value
–
<imped>
For TEM type connectors: reference impedance in
V
(without unit),
For WGUide type connectors: cutoff frequency in Hz (without unit)
Range [def. unit]
Ref. impedance: 1
µ
to 1000 M
V
Cutoff frequency: 0 Hz to 1000 GHz
*RST value
–
SCPI, Command
Types
Device-specific, command or query
Example:
CORR:CONN 'USERCON',TEM,GEND,1.00000,50
Define a TEM type connector type names USERCON.
CORR:CONN? 'USERCON'
Query the properties of the configured connector type.
CORR:CONN:DEL 'USERCON'
Delete the configured connector type.
[SENSe<Ch>:]CORRection:CONNection:DELete '<conn_name>'
Deletes a user-defined connector type named '<conn_name>'.
<Ch>
Channel number