R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
256
Remote
control:
The commands in the
[SENSe<Chn>:]SEGMent...
subsystem define all sweep
segment settings.
Inserting and Deleting Segments
Three buttons on the left side below the table in the
Define Segments
dialog extend or shorten the
segment list:
•
Insert New Segment
adds a new segment to the list. The new segment is inserted after the active segment. The
segment numbers (#) of all segments after the new segment are adapted.
To avoid overlapping segments, the new segment ranges from the
Stop
value of the previously
active segment to either the
Start
valueof the next segment or the maximum frequency of the
analyzer. The analyzer places no restriction on the number of segments in a sweep range.
•
Delete Segment removes the selected segment from the list.
•
Del All Segments clears the entire segment list so it is possible to define or load a new segmented
sweep range.
Remote control:
[SENSe<Chn>:]SEGMent<Seg>:INSert
[SENSe<Chn>:]SEGMent<Seg>:DELete
[SENSe<Chn>:]SEGMent<Seg>:DELete:ALL
Individual Settings
The options in the
Individual Settings
panel can be used to vary the channel settings of every individual
segment in the list.
The first sweep segment is created with the channel settings defined for general sweep types. When any
further sweep segment created, it uses the channel settings of the previously active segment. Each
selected (checked) option adds a column to the segment list.
•
Name adds a column to assign a name to each segment. A segment name is a string that may
contain letters, numbers and special characters.
•
Power defines the internal source Power for each individual sweep segment.
•
Meas Bandwidth defines the Meas Bandwidth for each individual sweep segment.
•
Filter Type defines the Filter Type(selectivity) of the IF filter used for each sweep segment.
•
MeasDelay
sets a delay time allowing the DUT to settle before the hardware settings of the
analyzer are changed and a new partial measurement is started.
This setting is not available on
R&S ZVL13 analyzers.
•
Attenuator 1/2 select the attenuation factors for the received waves b1 and b2, respectively.
Remote control:
[SENSe<Chn>:]SEGMent<Seg>:POWer[:LEVel]
[SENSe<Chn>:]SEGMent<Seg>:BWIDth[RESolution]
[SENSe<Chn>:]SEGMent<Seg>:BWIDth[RESolution]:SELect