R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
487
'<Conn_Name>','<Ckit_Name>','<Std_No>',<Min_Freq>,<Max_Freq>,<El_Length>,<
Loss>,<Z0>,<C0> , <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>[, OPEN | SHORt |
MATCh, <Resistance>, <Port_1>, <Port_2>]
Defines the parameters of a non-ideal 1 port or 2-port calibration standard <std_type>. A particular
physical standard can be selected by specifying the name of the calibration kit and its serial number.
Depending on the standard type, only a subset of the parameters may be used; see table below.
<Ch>
Channel number. This suffix is ignored because calibration kits are channel-
independent.
<std_type>
Standard type. For reflection standards, the first character denotes the gender, e.g.:
FOPEN, MOPEN
: Open (f) or Open (m) standard.
The following reflection standards are supported:
MOPen, FOPen, MSHort,
FSHort, OSHort, MOSHort, FOSHort, MMTCh, FFTCh, MREFLect,
FREFLect
For transmission standards, the first two characters denotes the genders on both
ends, e.g.:
FFSNetwork, MFSNetwork, MMSNetwork:
Symm. network (ff), symm. network
(mf) or symm. network (mm) standard.
For a complete list of standard types refer to the table of standard types below.
Parameter
list
String parameters to specify the configured standard
('<Conn_Name>','<Ckit_Name>','<Std_No>') and numeric parameters defining its
properties. See detailed description below.
The following transmission standards have only 8 parameters (until <Z0>):
MMTHrough, MFTHrough, FFTHrough, MMLine, MFLine, FFLine
The following reflection and transmission standards have only 5 parameters
(until <Max_Freq>):
MSMatch, FSMatch, MMATten, MFATten, FFATten
*RST value
–
SCPI,
Command
Types
Device-specific, command or query (see examples below)
Example:
CORR:CKIT:FOP 'N 50 Ohm','Test Kit',...
Define the properties of the open (f) standard for a N 50
V
connector type in cal kit
'Test Kit'. See also example for
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
.
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the open (f) standard for a N 50
V
connector type in the
active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','Test Kit'
Query the properties of the open (f) standard for a N 50
V
connector type in cal kit
'Test Kit'.