R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
519
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and
thus the
Fast
filter type.
SEGM:BWID:SEL NORM
Activate an IF filter with high selectivity.
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol <Boolean>
Qualifies whether or not the
Filter Type
(selectivity) can be set independently for each sweep segment.
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Boolean>
ON – The filter type can be set independently for each sweep segment.
OFF – The filter type in all sweep segments is equal to the filter type for
unsegmented sweeps set via
SENSe<Ch>:BWIDth[:RESolution]:SELect
.
*RST value
OFF. The parameter is automatically switched to ON when a bandwidth is entered
using
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect
.
SCPI,
Command
Types
Device-specific, command or query
Example:
*RST; SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus
the
Fast
filter type.
SEGM:BWID:SEL NORM
Change the filter type to
Normal
.
BWID:SEL?
Query the (default) filter type for unsegmented sweeps. The response is
FAST
.
SEGM:BWID:SEL:CONT OFF
Couple the filter type in all segments and reset the filter type in segment no. 1 to the
unsegmented value
FAST
..
[SENSe<Ch>:]SEGMent:CLEar
Deletes all sweep segments in the channel. The command is equivalent to
[SENSe<Ch>:]SEGMent:DELet:ALL
.
[SENSe<Ch>:]SEGMent<Seg>:COUNt?
Returns the number of sweep segments in the channel including all segments that are switched off
(
[SENSe<Ch>:]SEGMent<Seg>[:STATe] OFF)
.
<Ch>
Channel number.