R&S ZVL
GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
168
•
Bandpass / Bandstop Absolute Level:
The bandpass / bandstop is the tallest/lowest peak in the
search range. To be valid, the peak must be above / below –x dB, where x is numerically equal to
the <x dB Bandwidth> value. The Lower Band Edge and Upper Band Edge values are given by
the frequencies where the trace is equal to –x dB.
Tracking
is a toggle function: Selecting the function repeatedly switches the tracking mode on and off.
Remote control:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute BFILter
CALCulate<Chn>:MARKer<Mk>:SEARch:TRACking
Level
Opens the numeric entry bar for the minimum excursion of the bandpass and bandstop peaks.
•
A bandpass peak must fall off on both sides by the specified
x dB Bandwidth
to be considered a
valid peak.
•
A bandstop peak must be
x dB Bandwidth
below the maximum level in the search range
(bandpass value) to be considered a valid peak.
Remote control:
CALCulate<Chn>:MARKer<Mk>:BWIDth <Level>
Search Range...
Opens the
Search Range Dialog
to confine the bandfilter search to a subrange of the sweep.
It is possible to define and store up to ten different search ranges for each trace. The bandfilter search is
performed using the markers
Mkr 1, ..., Mkr 4,
irrespective of the selected search range.
•
Search Range
Selects a search range for the bandfilter search. Full Span means that the search
range is equal to the sweep range. Besides, it is possible to store up to 10 customized search
ranges.
•
Start
defines the beginning of the search range. Start must be smaller than the Stop value,
otherwise the search will not be initiated.
•
Stop
defines the end of the search range. Stop must be larger than the Start value, otherwise the
search will not be initiated.
•
Range Limit Lines On
displays two vertical lines indicating the Start and the Stop value of the
current bandfilter search range in the diagram area. This function is enabled as soon as one of the