1.9.1 Operation with and without Administrator Rights
With firmware version V3.xx (image version 3.11) and higher, the network analyzer
may be operated with or without administrator rights. Some administrative tasks (e.g. a
firmware update) do require administrator rights; refer to the detailed description of the
SETUP softkeys in chapter 4 of the operating manual.
In the default configuration, auto login is enabled, and an account with administrator
rights is active. This means that no password is required, and the full functionality of
the analyzer is available to any user. An additional user account without administrator
rights is pre-defined.
To install new fimware using the update tool,
To change the user account, access Windows XP's Start menu, e.g. by pressing
SETUP – General Setup – More – Open Start Menu
(see section Accessing Windows
XP's Start Menu on p.
37
) and click
Settings – Control Panel – User Accounts
. In the
dialog opened, you can also change the passwords for the user accounts. User
accounts with the following user names and passwords are pre-defined::
Description User
name
Pasword
Default account, with administrator rights
Account without administrator rights
NormalUser
894129
(De)activating auto login
After being turned off, the R&S ZVB will return to the
Instrument
account without
prompting for a password. To de-activate the automatic login mechanism,
1. Access the Start menu and select
Run...
2. In the dialog box opened, enter C:\R_S\INSTR\USER\NO_AUTOLOGIN.REG and
confirm.
Auto login is deactivated. The next time you switch on the R&S ZVL, you will have to
enter a user name and a password. Proceed in the same way, using the
C:\R_S\INSTR\USER\AUTOLOGIN.REG file, to re-activate automatic login.
Firmware upgrade and downgrade
An update from a firmware version <V3.xx to version V3.xx or higher does not replace
the image. To replace the image version, contact your R&S service representative. The
image upgrade is possible on instruments with a DVI-D connector on the rear panel.
A downgrade of the firmware from V3.xx or higher to version <V3.xx requires the
BackGrade
program, which can be accessed from the Windows XP Start menu of the
analyzer:
1. Access the Start menu.
2. Click
Programs – Accessories – BackGrade
and confirm with OK.
3. Perform the firmware update using the
Instrument Update Tool
described above.