R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
379
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STARt <numeric_value>
Defines the start value of the search range selected with
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
<numeric_value>
Beginning of the search range.
Range [def. unit]
Maximum allowed sweep range, depending on the instrument model and on
the sweep type. [Hz, dBm or s, depending on the sweep type]
*RST value
0 Hz
SCPI, Command
Types
Device-specific, command or query.
Example:
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER:STOP <numeric_value>
Defines the stop value of the search range selected with
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
.
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
<numeric_value>
End of the search range.
Range [def. unit]
Maximum allowed sweep range, depending on the instrument model and on
the sweep type. [Hz, dBm or s, depending on the sweep type]
*RST value
0 Hz
SCPI, Command
Types
Device-specific, command or query.
Example:
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER <numeric_value>
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum | MINimum | RPEak
| LPEak | NPEak | TARGet | LTARget | RTARget | BFILter
Selects a search mode for marker no. <Mk> and initiates the search. The marker must be created before
using
CALCulate<Chn>:MARKer<Mk>[:STATe] ON
(exception: bandfilter search).
<Chn>
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. For a bandfilter search (
BFILter
) this
numeric suffix is ignored and may be set to any value because the bandfilter
search functions always use markers Mkr1 to Mkr4.
Parameters
See list of parameters below.
*RST value
–
SCPI,
Command
Types
Device-specific, no query.
Example: