R&S ZVL
Interfaces and Connectors
Rear Panel Connectors
Operating Manual 1303.6580.32-05
585
EXT REF
BNC female connector, used as an input or output for the 10 MHz reference clock signal.
The function of the
10 MHz REF
connector depends on the
Int. Reference
or
Ext. Reference
setting in the
SETUP
menu:
•
If
Int. Reference
is active,
10 MHz REF
is used as an output connector for the
10 MHz internal
reference clock signal of the analyzer. This configuration requires the OCXO option (R&S FSL–
B4, OCXO Reference Frequency).
•
If
Ext. Reference
is active,
10 MHz REF
is used as an input connector for an external 10 MHz
reference clock signal. The external reference signal must meet the specifications of the data
sheet. The internal reference signal is synchronized to the external signal.
The AC power switch also interrupts the power supply of the OCXO. When you switch the
instrument back on, be sure to comply with the extended warm–up phase specified in the
data sheet.
POWER SENSOR (Option R&S FSL–B5, Additional Interfaces)
LEMOSA female connector, used for connecting power sensors of the R&S NRP–Zxy family. With an
adapter cable R&S NRP–Z4, the USB port on the front panel can be used for this purpose.
NOISE SOURCE CONTROL (Option R&S FSL–B5, Additional
Interfaces)
BNC female connector, provides the supply voltage for an external noise source, e.g. to measure the
noise figure and gain of amplifiers and frequency converting DUTs.
Conventional noise sources require a voltage of +28 V in order to be switched on and 0 V to be switched
off. The output supports a maximum load of 100 mA.