235
Remote
control:
CALCulate<Ch>:TRANsform:VNETwork:SENDed...
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<Ph_pt>
'<file_name>'
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<Ph_pt> '<file_name>'
Cal Manager
Opens a dialog to store system error correction and power correction data to the cal pool and to assign
stored correction data to channels.
Cal pool
The cal pool is a collection of correction data sets (cal groups) that the analyzer stores in a common
directory
C:\Rohde&Schwarz\NWA\Calibration\Data.
Cal groups in the pool can be applied to different
channels and setups. Each cal group is stored in a separate file named <CalGroup_name>.cal. The cal
group name can be changed in the
Calibration Manager
dialog.
The
Calibration Manager
dialog contains three main panels:
•
Calibration State
lists all channels and their current calibration state.
•
Calibration Pool
contains a list of all stored correction data sets.
•
Calibration Properties
displays the basic channel settings, the
System Error Correction Data,
and