R&S ZVL
GUI Reference
SETUP Key
Operating Manual 1303.6580.32-06
301
GPIB Address
Opens an edit dialog box to enter the GPIB address. Values from 0 to 30 are allowed. The default address
is 20.
If, e.g. after a firmware update, the R&S ZVL does not maintain the GPIB address after reboot, the
shutdown file needs to be recreated. Perform the following steps:
Set the GPIB address.
Create a shutdown file by switching the R&S FSL in the standby mode:
Press the ON/STANDBY key on the front panel and wait until the yellow LED is on. With the battery
pack option, use a USB keyboard and terminate the R&S ZV firmware with ALT+F4 to create the
shutdown file.
If the shutdown file was created once, the R&S ZVL boots with exactly those settings after a reboot with
the main switch on the rear panel. During a firmware update or cold boot for service reasons the shutdown
file is deleted, and the setup is necessary once again.
Remote Control:
SYST:COMM:GPIB:ADDR 20
ID String Factory
Selects the default response to the *IDN? query.
ID String User
Opens an edit dialog box to enter a user–defined response to the *IDN? query. Max. 36 characters are
allowed.
GPIB Language
Only the remote control language SCPI is available for the R&S ZVL and is set by default.
Display Update On/Off
Defines whether the instrument display is switched off when changing from manual operation to remote
control. In remote control mode, this softkey is displayed in the local menu.
Remote Control:
SYST:DISP:UPD
GPIB Terminator LFEOI/EOI
Changes the GPIB receive terminator.
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary data transfers (e.g.
trace data) from the control computer to the instrument, the binary code used for <LF> might be included
in the binary data block, and therefore should not be interpreted as a terminator in this particular case.
This can be avoided by changing the receive terminator to EOI.
Remote Control:
SYST:COMM:GPIB:RTER