R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
493
two separate calibrations are performed at ports 1 and 2.
*RST value
ON
AUTO |
<delay |
phase>
Optional entry of delay time or phase for UTHRough standard:
AUTO – The analyzer determines the delay time or phase during the calibration sweep
<delay or phase> – entry of the delay time in ps (for non-dispersive standards) or of an
estimate of the phase at the start frequency of the sweep in deg (for dispersive
standards). If an estimate of the start phase is entered, the analyzer uses the
calculated
value which is closest to the estimate.
*RST value
AUTO
SCPI,
Command
Types
Confirmed (with device-specific standards), no query
Example:
See
[SENSe<Ch>:]
CORRection:COLLect:SAVE.
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe <Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the standards is stored
after the calibration is completed. The setting is valid for the current calibration, where it overwrites the
global setting (
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault
). A new
calibration deletes the calibration data acquired in previous calibrations.
<Ch>
Channel number of the calibrated channel.
<Boolean>
ON | OFF - Keep measurement data on or off.
*RST value
OFF
SCPI, Command
Types
Device-specific, command or query
Example:
CORR:COLL:RSAV:DEF ON
Generally keep the raw measurement data of the standards after the
calibration is completed.
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as calibration
type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the raw measurement
results of this standard.
CORR:COLL:RSAV OFF
To save disk space, delete the current raw calibration data after the calibration
is completed.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply them to the active
channel.