![Rohde & Schwarz ZVL series Operating Manual Download Page 149](http://html.mh-extra.com/html/rohde-and-schwarz/zvl-series/zvl-series_operating-manual_1477904149.webp)
R&S ZVL
GUI Reference
Trace Menu
Operating Manual 1303.6580.32-06
134
•
Selecting one of the existing channel names assigns the current trace to the existing channel.
•
Selecting
New
creates a new channel and assigns the current trace to the new channel. The new
channel is named Ch <n>, where <n> is the largest of all existing channel numbers plus one. The
name can be changed in the Channel Manager.
Remote control:
CALCulate<Ch>:PARameter:SDEFine <Trace Name>, < Meas Parameter>
Trace Manager
Opens a dialog to perform the actions in the
Trace Select
menu systematically for all traces and diagram
areas.
All
existing traces of the current setup are listed in a table with several editable (white) or non-editable
(gray) columns. Below the table the
Trace Manager
provides the following buttons:
•
Add/Delete...
opens a dialog to add a new trace or delete a trace.
•
Coupling
opens a dialog to define coupling criteria (channel, scale) for all traces in the table.
•
Sort Table
opens a dialog to change the order of the traces (rows) in the table.
Columns in the Trace Manager table
•
Name
indicates the current trace name. The default names for new traces are Trc<n> where <n>
is a current number. Current numbers in the trace names are necessary to make automatic
assignments, e.g. decouple the channel settings in the
Coupling
dialog.
•
On
indicates whether the trace is displayed on the screen
(On)
or invisible.
•
Meas
indicates the measured parameter.
•
Type
indicates whether the trace is a data trace
(DAT),
displaying the current measurement data,
or a memory trace
(MEM)
.
•
Channel
indicates the channel of each trace.
•
Area
indicates the diagram area of each trace.
•
Scale
shows which traces use common scaling and format settings.
Rules for trace names
For trace names, the following restrictions apply: