R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
505
[SENSe<Ch>:]CORRection:COLLect:METHod
on port numbers)
One-port normalization
(reflection) using an open
standard
REFL1
REFL2
REFL12
'SCORR3'
'SCORR9'
'SCORR3' and 'SCORR9'
Full one port
FOPort1
FOPort2
FOPort12
'SCORR1' to 'SCORR3'
'SCORR7' to 'SCORR9'
'SCORR1' to 'SCORR3' and
'SCORR7' to 'SCORR9'
Two-port normalization
FTRans
RTRans
FRTRans
'SCORR6'
'SCORR12'
'SCORR6' and 'SCORR12'
One path two port
FOPTport
ROPTport
'SCORR1' to 'SCORR3', 'SCORR6'
'SCORR7' to 'SCORR9',
'SCORR12'
TOSM
TOSM
'SCORR1' to 'SCORR12'
(at present the isolation terms
'SCORR4' and 'SCORR10' are not
included)
[SENSe<Ch>:]CORRection:DATA:PARameter?
Returns the sweep settings of the active system error correction for channel <Ch>.
<Ch>
Channel number of the calibrated channel.
Response
Five values, corresponding to the start and stop frequency of the calibration
sweep, the number of points, the source power, and the sweep type (LIN | LOG |
SEGM).
*RST value
– (the command generates an execution error if no system error correction is
active)
SCPI,
Command
Types
Device-specific, query only
Example:
See [SENSe<Ch>:]CORRection:DATE?
[SENSe<Ch>:]CORRection:DATE?
Returns the date and time when the active system error correction data for channel <Ch> was acquired.
<Ch>
Channel number of the calibrated channel.
*RST value
– (the command generates an execution error if no system error correction is active)
SCPI,
Command
Types
Device-specific, query only
Example:
CORR:COLL:METH REFL1
Select a one-port normalization at port 1 as calibration type.
CORR:COLL OPEN1
Measure an open standard connected to port 1 and store the measurement results
of this standard.
CORR:COLL:SAVE