R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
494
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:RSAVe:DEFault <Boolean>
Activates or deactivates the calibration mode where the raw measurement data of the standards is stored
after the calibration is completed. The setting remains valid for all subsequent calibrations until it is
changed explicitly. A new calibration deletes the calibration data acquired in previous calibrations.
<Ch>
Channel number of the calibrated channel.
<Boolean>
ON | OFF - Keep measurement data on or off.
*RST value
OFF
SCPI, Command Types
Device-specific, command or query
Example:
See
[SENSe<Ch>:]CORRection:COLLect[:AQUire]:RSAVe.
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected THRough | OPEN |
SHORt | MATCh | NET | ATT | REFL | SLIDE | LINE1 | LINE2,<port_no>,<port_no>
Starts a calibration measurement in order to acquire measurement data for the selected standards. The
standards are reflection or transmission standards and can be connected to arbitrary analyzer ports.
<Ch>
Channel number of the calibrated channel.
THRough ...
LINE2
Standard types:
Through, Open, Short, Match, Symmetric Network (NET),
Attenuation (ATT), Reflect, Sliding Match (SLIDe), Line1 (LINE1 and LINE are
synonymous), Line2
.
<port_no>
Port numbers of the analyzer. For a transmission standard (through, line,
attenuation, symmetric network) the input and output port numbers must be
specified, for reflection standards, only one port number is required.
*RST value
–
AUTO |
<delay |
phase>
Optional entry of delay time or phase for UTHRough standard:
AUTO – The analyzer determines the delay time or phase during the calibration
sweep
<delay or phase> – entry of the delay time in ps (for non-dispersive standards) or of
the phase at the start frequency of the sweep in deg (for dispersive standards)
*RST value
AUTO
SCPI,
Command
Types
Device-specific, no query
Example:
See [SENSe<Ch>:]CORRection:COLLect:SAVE:SELected.
[SENSe<Ch>:]CORRection:COLLect:AUTO '<file_name>',<port_no>{,<port_no>}
Selects and initiates a one-port or two-port automatic calibration
at arbitrary analyzer and cal unit ports.
If the test setup contains a high attenuation the analyzer may fail to detect the cal unit ports
connected to each of its ports. In this case use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
.
If several calibration units are connected, use
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
to