R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
518
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and
thus 10 kHz measurement bandwidth.
SEGM:BWID 1 MHZ
Increase the resolution bandwidth to 1 MHz.
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol <Boolean>
Qualifies whether or not the
Meas Bandwidth
can be set independently for each sweep segment.
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Boolean>
ON: The bandwidth can be set independently for each sweep segment.
OFF: The bandwidth in all sweep segments is equal to the bandwidth for
unsegmented sweeps set via
SENSe<Ch>:BWIDth[:RESolution]
.
*RST value
OFF. The parameter is automatically switched to ON when a bandwidth is entered
using
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
.
SCPI,
Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus
10 kHz measurement bandwidth.
SEGM:BWID 1 MHZ
Increase the resolution bandwidth to 1 MHz.
SEGM:BWID:CONT OFF
Couple the bandwidths in all segments and reset the bandwidth in segment no. 1 to
the initial value.
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect FAST | NORMal
Defines the
Filter Type
(selectivity) in sweep segment no. <Seg>. At the same time, the command
activates separate filter type settings in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]SELect:CONTrol ON
).
<Ch>
Channel number.
<Seg>
Sweep segment number.
FAST
Use IF filter with normal selectivity and short settling time.
NORMal
Use IF filter with higher selectivity but larger settling time.
*RST value
FAST
SCPI, Command
Types
Device-specific, command or query