R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
252
The
Sweep
menu provides the following settings:
•
Sweep Type
defines the position of sweep points (frequency values) in the sweep range.
•
Number of Points
sets the total number of measurement points over the sweep range.
•
Frequency Step Size
sets the distance between two consecutive frequency sweep points.
•
Meas Delay
delays the start of each sweep.
•
Trigger
selects the trigger mode for the measurement.
•
Restart
aborts the current measurement and restarts a new sweep sequence.
•
If
Single (All Chans)
is selected the measurement is terminated after a single sweep or a group of
single sweeps defined in the
Define Restart
dialog.
•
Define Restart
opens a dialog to specify which channels are affected and how many sweeps are
repeated.
•
Average On
activates or de-activates the sweep average. With average on the measurement
results are averaged over a selected number of consecutive sweeps (Average Factor).
•
Average Factor
defines the number of consecutive sweeps to be averaged.
•
Restart Average
starts a new average cycle, clearing all previous results and thus eliminating their
effect on the new cycle. The new cycle is started as fast as possible; an ongoing sweep is
terminated immediately.
Sweep Type
The
Sweep Type
submenu defines the frequency sweep type and the position of the sweep points across
the sweep range.
•
Lin Frequency
is the default sweep type. The stimulus frequency is swept in equidistant steps