Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
482
Turn off the RF source power.
PROGram...
This subsystem controls external application programs that can be run on the analyzer.
PROGram[:SELected]:EXECute '<file_name>[ <command>]'
Starts an application program or open a file using an application available on the analyzer.
'<file_name>'
Name and path of an application program to be executed or of a file to be opened.
The path can be defined as an absolute path (e.g. 'c:\...') or relative to the current
directory (MMEMory:CDIRectory). Blanks in the file name are not allowed because
everything after a blank is interpreted as a <command>.
<command>
Command name in the selected application <file_name> (optional).
*RST value
–
SCPI,
Command
Types
Confirmed, no query.
Example:
PROG:SEL:NAME PROG
Selects general program execution on the analyzer.
MMEM:CDIR 'C:\Program Files\Rohde&Schwarz\Network Analyzer\Bin'
Go to the program directory.
PROG:EXEC 'iecwin32.exe'
Start the application iecwin32.
PROG:EXEC:WAIT
Lock command execution and manual control of the analyzer until iecwin32 is
paused or stopped.
PROGram[:SELected]:NAME PROG
Selects the application to be run on the analyzer. At present, only the general parameter PROG is
available:
PROGram[:SELected]:EXECute
can start any program.
Use this command in order to avoid problems should the default value change in future firmware
versions.
PROG
Any program running under Windows XP or any file that can be opened with an
application program available on the analyzer.