R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
259
•
If the sweep range is defined by means of the
Start
and
Stop
variables, both the
Stop
value and
the
Number of Points
can vary as the
Stimulus Step Size
is changed. The
Stop
value is changed
as little as possible so that the condition
Stimulus Step Size = (Stop – Start) / (Number of Points –
1)
can be fulfilled
.
Changing the
Start and Stop
values
modifies the
Stimulus Step Size.
•
If the sweep range is defined by means of the
Center
and
Span
variables, both the
Span
value
and the
Number of Points
can vary as the
Stimulus Step Size
is changed. The
Span
is reduced as
little as possible so that the condition
Stimulus Step Size = (Stop – Start) / (Number of Points – 1)
can be fulfilled
.
Changing the
Span
modifies the
Stimulus Step Size.
This setting is valid for linear frequency sweeps only. It does not apply to logarithmic and segmented
sweeps. Increasing the Frequency Step Size generally increases the measurement time.
Remote control:
[SENSe<Ch>:]SWEep:STEP
Meas Delay
Sets a delay time before the start of each partial measurement. This setting is not available on R&S
ZVL13 network analyzers.
A measurement delay time increases the accuracy, in particular in measurements on DUTs with long
settling times (e.g. quartzes, SAW filters).
Remote control:
[SENSe<Ch>:]SWEep:DWELl
Restart
Stops the current measurement and restarts a measurement sequence. In
Single
sweep mode a new
single sweep sequence is started.
Remote control:
INITiate<Ch>[:IMMediate]
Single (All Chans)
Toggles between single sweep and continuous sweep mode.
•
In continuous mode
(Single (All Chans)
not selected), the analyzer measures continuously,
repeating the current sweep.
•
In single sweep mode, the measurement is stopped after the number of sweeps selected in the
Define Restart
dialog.
Press
Single
to activate the single sweep mode and start a single sweep sequence. To start further
sweep sequences press
Restart.
Remote control:
INITiate<Ch>:CONTinuous ON | OFF