R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
483
*RST value
PROG
SCPI, Command
Types
Confirmed, command or query.
Example:
See
PROGram[:SELected]:EXECute.
PROGram[:SELected]:WAIT
Locks command execution and manual control of the analyzer while a program started via
PROGram[:SELected]:EXECute
is running. The analyzer does not execute any further commands or
queries until the program is stopped or paused.
Parameters
None.
*RST value
–
SCPI, Command
Types
Confirmed, command or query. The query also locks command execution; it
returns 1 when the executed program is stopped or paused.
Example:
See
PROGram[:SELected]:EXECute.
[SENSe...]
[SENSe<Ch>:]AVERage...
This subsystem sets sweep averaging parameters. The sweep average is a noise-reduction technique
which consists of calculating each measurement point as an average of the same measurement point over
several consecutive sweeps.
[SENSe<Ch>:]AVERage:CLEar
Starts a new average cycle, clearing all previous results and thus eliminating their effect on the new cycle.
<Ch>
Channel number. If unspecified the numeric suffix is set to 1.
*RST value
–
SCPI, Command Types
Confirmed, no query
Example:
AVER:COUN 15; AVER ON
Set the average factor for channel 1 to 15 and enable the sweep average.
AVER:COUN 5; CLE
Reduce the average factor and restart the average.