Create a setup named 'Setup_2' and make it the active
setup.
MEM:SEL 'Set1'
Activate the default setup Set1.
M
MEM:STOR:STAT 1, 'C:\R_S\Instr\user\Nwa\RecallSets\Set1.nwa'; MEM:DEL
'Set1.nwa
'
Store the active setup Set1 to a file, renaming it Set1.nwa. Close the setup.
MMEMory...
The MMEMory system provides mass storage capabilities for the analyzer.
Internal and external mass storage
The mass storage of the analyzer may be internal or external. The internal mass storage device can be
any section of the internal hard disk (mapped to drive C:\). The external mass storage device can be a
floppy disk inserted into the drive at the front panel of the instrument, which is mapped to drive a:\ (see
MMEMory:MSIS
), a USB memory stick connected to one of the USB ports (mapped to any free drive
letter) or a network connection.
File and directory names
The <file_name> and <directory_name> parameters are strings. Some commands use a fixed directory;
for others the <file_name> can contain the complete path including the drive name and all subdirectories,
e.g. 'C:\TEMP\TRASH\test.txt' for the file named test.txt in the TEMP\TRASH subdirectory of the internal
hard disk drive C:\. If no complete path is specified, the file location is relative to the current directory
(queried with
MMEMory:CDIRectory?
). The file name itself may contain the period as a separator for
extensions.
File and directory names can be chosen according to Windows™ conventions; the restrictions placed on
file names known from DOS systems do not apply. All letters and numbers are allowed, as well as the
special characters "_", "^", "$", "~", "!", "#", "%", "&", "-", "{", "}", "(", ")", "@" and "`". Reserved file names
are CON, AUX, COM1, ..., COM4, LPT1, ..., LPT3, NUL and PRN.
The use of wildcards ? and * is not allowed.
MMEMory:AKAL:FACTory:CONVersion '<directory_name>'
Converts the calibration data of the standards in the active calibration unit
(
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
) to Touchstone format and copies it to the specified
directory.
'<directory_name>'
String parameter to specify the directory.
*RST value
–
SCPI, Command
Types
Device-specific, command or query (returns the current directory).
Example:
MMEM:AKAL:FACTory:CONVersion 'C:\AKAL\Touchstone'
Convert and copy the calibration data of the standards