R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
525
<Unused>
Ignored parameter, should be set to the default value 0.
<Meas.
Bandwidth>
Resolution bandwidth in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]
.
Range [def. unit]
1.0E-6 Hz to 5 MHz [Hz].
SCPI,
Command
Types
Device-specific, no query.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the second segment
will be between 1.5 MHz and the maximum frequency of the analyzer.
[SENSe<Ch>:]SEGMent<Seg>:OVERlap <Boolean>
Queries whether the analyzer supports overlapping sweep segments.
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Boolean>
ON | OFF: No effect.
*RST value
OFF. If used as a query, the command returns the information that overlapping
sweep segments are not supported
(OFF)
.
SCPI, Command
Types
Device-specific, command (no effect) or query
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel] <Power>
Defines the
Power
of the internal signal source in sweep segment no. <Seg>. At the same time, the
command activates separate power control in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:POWer[:Level]:CONTrol ON
).
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Power>
Internal source power.
Range [def. unit]
–40 dBm to +10 dBm. The exact range depends on the analyzer model; refer to
the data sheet [dBm].
*RST value
–10 dBm.
SCPI, Command
Types
Device-specific, command or query
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and
thus –10 dBm internal source power.
SEGM:POW -20