R&S ZVL
GUI Reference
SETUP Key
Operating Manual 1303.6580.32-06
290
Service
g
Reset Password
Selftest
Selftest Results
Password
Service Function
Reference Int/Ext
Switches between the internal and external reference signal source. The default setting is internal
reference. It is important, that the external reference signal is deactivated when switching from external to
internal reference to avoid interactions.
If the reference signal is missing when switching to external reference, the message
EXREF
is displayed
to indicate that no synchronization is possible.
The R&S ZVL can use the internal reference source or an external reference source as frequency
standard from which all internal oscillators are derived. A 10 MHz crystal oscillator is used as internal
reference source. In the external reference setting, all internal oscillators of the R&S ZVL are
synchronized to the external reference frequency (also 10 MHz).
Remote Control:
ROSC:SOUR INT
Firmware Update
Opens a submenu to install firmware versions. The installation of a new firmware version can be
performed via USB or LAN interface; see
Firmware Update
. With firmware versions V3.00 and higher, a
user account with administrator rights is necessary to install a printer (see
Operation with and without
Administrator Rights
).
Firmware Update
Opens the
Firmware Update
dialog box. The update path is changed by entering the new path or via the
Browse
button. The installation is started via the
Execute
button; see
Firmware Update
.
Remote Control:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Option Licenses
Opens a submenu to install options. With firmware versions V3.00 and higher, a user account with
administrator rights is necessary to install a printer (see
Operation with and without Administrator Rights
).