R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
517
Types
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the second segment
will be between 1.5 MHz and the maximum frequency of the analyzer.
[SENSe<Ch>:]SEGMent<Seg>:ATTenuation<port_no> <Attenuation>
Defines the attenuation factors for the received waves b
1
and b
2
in segment no. <Seg>. .
<Ch>
Channel number.
<Seg>
Sweep segment number.
<port_no>
Test port number of the analyzer, 1 (wave b
1
) or 2 (wave b
2
). If unspecified the
numeric suffix is set to 1.
<Attenuation>
Attenuation factor for the received wave.
Range [def. unit]
0 dB to +30 dB. UP and DOWN increment/decrement the attenuation in 5 dB
steps. The analyzer rounds any entered value below the maximum attenuation
to the closest step.
*RST value
+10 dB
SCPI, Command
Types
Device-specific, with query.
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0, 10KHZ
Create a sweep segment with a sweep range between 1.0 MHz and 1.5 MHz.
SEGM:ATT1 5; ATT2 15
Define an input attenuation of 5 dB for the b
1
wave, 15 dB for the b
2
wave.
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution] <Bandwidth>
Defines the resolution bandwidth of the analyzer
(Meas Bandwidth)
in sweep segment no. <Seg>. At the
same time, the command activates separate bandwidth setting in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:CONTrol ON
).
<Ch>
Channel number.
<Seg>
Sweep segment number.
<Bandwidth>
Internal source power.
Range [def. unit]
1.0E-6 Hz to 5 MHz [Hz]. UP and DOWN increment/decrement the bandwidth in
1-2-5 steps for each decade. The analyzer rounds up any entered value between
these steps and rounds down values exceeding the maximum bandwidth.
*RST value
10 kHz
SCPI,
Command
Types
Device-specific, command or query