R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
366
Parameters
Keyword for the physical unit of the response values.
Range [def. unit]
The parameters form four groups:
S and SINV select relative units (dB) for the limit line.
Y selects admittance units (S/Siemens).
Z selects impedance units (
V
).
YREL and ZREL select dimensionless numbers (U).
[–]
*RST value
–
SCPI, Command Types
Device-specific, no query.
CALCulate<Chn>:LIMit:RDOMain:FORMat COMPlex | MAGNitude | PHASe | REAL |
IMAGinary | SWR | GDELay | L | C
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via
CALCulate<Chn>:LIMit:CONTrol:DOMain.
This command is complemented by
CALCulate<Chn>:LIMit:RDOMain:COMPlex
and
CALCulate<Chn>:LIMit:RDOMain:SPACing
.
<Chn>
Channel number used to identify the active trace.
Parameters
Keyword for the physical unit of the response values.
Range [def. unit]
The parameters form four groups:
COMPlex, REAL, IMAGinary, and SWR select dimensionless
numbers (U) for the limit line.
MAGNitude selects relative units (dB).
PHASe selects phase units (deg).
GDELay selects time units (s).
L selects inductance units (H/Henry).
C selects capacitance units (F/Farad).
[–]
*RST value
–
SCPI, Command
Types
Device-specific, no query.
CALCulate<Chn>:LIMit:RDOMain:SPACing LINear | LOGarithmic | dB | SIC
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via
CALCulate<Chn>:LIMit:CONTrol:DOMain.
This command is complemented by
CALCulate<Chn>:LIMit:RDOMain:COMPlex
and
CALCulate<Chn>:LIMit:RDOMain:FORMat
.
<Chn>
Channel number used to identify the active trace.