R&S ZVL
GUI Reference
Channel Menu
Operating Manual 1303.6580.32-06
237
setup is already active, nothing is changed.
•
The calibrated setups are automatically stored in the
C:\R_S\Instr\user\Nwa\Calibration\RecallSets
directory. A message box pops up if the directory is
empty, e.g. because no calibration was performed yet.
Automatic Calibration (Procedure)
A
Calibration Unit
is an integrated solution for automatic system error correction of vector network
analyzers. For an overview of available types refer to section Automatic Calibration (Introduction). The use
of the calibration units is described below.
Attention!
Please observe the safety instructions in the "Technical Information"
provided with the calibration unit to avoid any damage to the unit and the
network analyzer. Safety-related aspects of the connection and
operation of the units are also reported in the sections below.
Connecting the Calibration Unit to the Network Analyzer
The calibration unit provides two N 75
V
(f) connectors.
To connect the unit...
Switch on and start-up your network analyzer.
To protect your equipment against ESD damage use the wrist strap and grounding cord supplied with
the instrument and connect yourself to ground. GND
Connect the USB type A connector of the USB cable to any of the USB type A connectors on the front
panel of the analyzer. Connect the USB type B connector of the USB cable to the USB type B
connector of the calibration unit. You can also connect the unit before switching on the analyzer.
Wait until the operating system has recognized and initialized the new hardware. When the unit is
connected for the first time, this may take longer than in normal use.
The unit is ready to be used, see
Performing an Automatic Calibration
below.