CALCulate<Chn>:STATistics[:STATe] <Boolean>
Displays or hides all statistical results in the diagram area of trace no. <Chn>.
You can display or hide the
Min/Max/Peak-Peak, Mean/Std Dev, RMS,
and the
Phase Delay/El
Length
results separately; see example below.
<Chn>
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Statistical info field on or off.
*RST value
OFF
SCPI, Command
Types
Device-specific, command or query.
Example:
*RST; CALC:STAT:MMPT ON
Reset the instrument, hiding all statistical results. Display the
Min/Max/Peak-
Peak
results.
CALC:STAT:MSTD ON
Display the
Mean/Std Dev
results in addition.
CALC:STAT:RMS ON
Display the
RMS
results in addition.
CALC:STAT:EPD ON
Display the
Phase Delay/El Length
results in addition.
CALC:STAT OFF
Hide all results.
CALCulate<Chn>:TRANsform...
This subsystem converts measured data from one representation to another and controls the
transformation into the time domain (with option ZVL-K3,
Time Domain)
.
CALCulate<Chn>:TRANsform:COMPlex S | Y | Z
Converts S-parameters into matched-circuit (converted) Y-parameters or Z-parameters and vice versa,
assuming that port no. i is terminated with Z
0i
so that the three parameter sets are equivalent and the
following formulas apply: