R&S ZVL
Command Reference
Instrument-Control Commands
Operating Manual 1303.6580.32-05
473
excludes them.
*RST value
ON
SCPI, Command
Types
Device-specific, command or query.
Example:
MMEM:SEL:HWS ON
Select the hardware settings.
MMEMory:SELect[:ITEM]:LINes:ALL <Boolean>
Adds all limit lines (activated and deactivated) to the list of device settings to be stored/loaded.
<Boolean>
ON | OFF – Includes the limit lines in the list of data subsets or excludes
them.
*RST value
ON
SCPI, Command
Types
Device-specific, no query.
Example:
MMEM:SEL:LIN:ALL ON
Select the limit lines.
MMEMory:SELect[:ITEM]:NONE
Deletes all data subsets from the list of device settings to be stored/loaded.
*RST value
–
SCPI, Command Types
Device-specific, no query.
Example:
MMEM:SEL:NONE
Exclude all data subsets.
MMEMory:SELect[:ITEM]:TRACe[:ACTive] <Boolean>
Adds the active traces to the list of data subsets of a save/recall device setting. Active traces are all
traces whose state is not blank.
<Boolean>
ON – Includes the active traces in the list of data subsets.
OFF – No traces will be stored.
*RST value
OFF
SCPI, Command Types
Device-specific, no query.
Example:
MMEM:SEL:TRAC ON
Select the active traces.